4 peak value test – Yokogawa PC-Based MX100 User Manual
Page 74
2-45
SM MX100-01E
Testing
1
2
3
4
5
6
7
2.10.3 ID Check
This test confirms whether the module is correctly identified when module information is
obtained from the MX100, MW100, or equivalent.
(If the OFF/ON cycle is not performed on the MX100, MW100, or equivalent, the ID
information is not updated. )
2.10.4 Peak Value Test
Wire the units as in the figure below, enter all settings (PWM output settings), then
confirm that the criteria are met.
Use an external power supply of 5 V, and load resistance of 5 Ω.
The wiring diagram in the figure below is an example of measurement on CH5. Actual
measurment should be performed on all channels.
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
DC
Power
supply
+
-
5Ω
DMM +
-
5V
V+
-
CH5
V+
-
CH6
V+
-
CH7
V+
-
CH8
+
-
Vext
+
-
Vext
V+
-
CH4
V+
-
CH3
V+
-
CH2
V+
-
CH1
Peak value check wiring diagram
PWM Outupt Settings
When using the MX100/DARWIN API for settings, enter settings that are equivalent to
the following.
Item
Setting
Remarks
Select output or no output Enabled
Select the source action
User output
Span
Need not specify
Output upon power ON
Preset value
Sets the output at power ON as a preset value.
Output upon error
Preset value
Sets the output upon error as a preset value.
Preset value
0.000 %
Sets 0% duty.
Pulse interval
10 ms
Sets the pulse interval to 10 ms.
Other settings
Not required
2.10 8-CH, Medium-Speed PWM Output Module