Ch1 ch2 ch3 ch4, Dip switch settings: 4-gauge method, Ndis terminal pin number – Yokogawa PC-Based MX100 User Manual
Page 59: Wiring diagram (ex.: ch2), 7 tests of the 4-ch strain input module

2-30
SM MX100-01E
Wiring for bridge voltage accuracy test (with -B12 and -B35)
Dip switch settings: 4-gauge method
ON
1 2 3 4 5
SW1 to SW4: OFF
SW5: ON
Wiring Diagram (Ex.: CH2)
Because the bridge voltage accuracy error will be equal the wiring resistance from the
clamp terminal to the load resistance, the effects of wiring resistance must be kept to
0.5% or less.
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
ON
1
2
3
4
5
ON
1
2
3
4
5
ON
1
2
3
4
5
ON
1
2
3
4
5
A/+V
B/ L
C/-V
D/ H
CH1
A/+V
B/ L
C/-V
D/ H
CH2
A/+V
B/ L
C/-V
D/ H
CH3
A/+V
B/ L
C/-V
D/ H
CH4
Load resistance
+
-
Waveform measuring
instrument
Wiring for bridge voltage accuracy test (with -NDI)
Ndis terminal pin number
A
G
B
C
F
E
D
Pin name
Signal name
A
B
C
D
E
F
G
P
L
N
H COM
PS NS
Wiring Diagram (Ex.: CH2)
CH1
CH2
CH3
CH4
A1002JC
Load resistance
+
-
P
N
PS
NS
Waveform measuring
instrument