2 withstanding voltage test, Procedure – Yokogawa PC-Based MX100 User Manual
Page 44
2-15
SM MX100-01E
Testing
1
2
3
4
5
6
7
2.5.1
Insulation Resistance Test
Attach the MX110 10-ch Medium-Speed Universal Input Module to the MX100, MW100,
or equivalent before performing this test.
Procedure
Wire the instrument as shown in the figure below, measure insulation resistance, then
confirm that the results meet the criteria below.
Measured Item
Criteria
Input terminal to earth terminal on the MX100, MW100, or
equivalent
20 MΩ or more (at DCV 500 V)
Between input terminals
20 MΩ or more (at DCV 500 V)
Between the input terminal and
earth terminal on the MX100 or equivalent
Between input terminals
+/A 1
–/B 2
b
3
1ch
+/A 1
–/B 2
b
3
2ch
+/A 1
–/B 2
b
3
3ch
+/A 1
–/B 2
b
3
4ch
+/A 1
–/B 2
b
3
+/A 1
–/B 2
b
3
+/A 1
–/B 2
b
3
+/A 1
–/B 2
b
3
+/A 1
–/B 2
b
3
+/A 1
–/B 2
b
3
5ch
6ch
7ch
8ch
9ch
10ch
Insulation resistance
tester or
Withstanding voltage
tester
+/A 1
–/B 2
b
3
1ch
+/A 1
–/B 2
b
3
2ch
+/A 1
–/B 2
b
3
3ch
+/A 1
–/B 2
b
3
4ch
+/A 1
–/B 2
b
3
+/A 1
–/B 2
b
3
+/A 1
–/B 2
b
3
+/A 1
–/B 2
b
3
+/A 1
–/B 2
b
3
+/A 1
–/B 2
b
3
5ch
6ch
7ch
8ch
9ch
10ch
Insulation resistance
tester or
Withstanding voltage
tester
2.5.2
Withstanding Voltage Test
Attach the 10-ch Medium-Speed Universal Input Module to the MX100, MW100, or
equivalent before performing this test.
Procedure
Wire the instrument in the same manner as in section 2.5.2, “Withstanding Voltage Test,”
then confirm that the results meet the criteria below.
With the withstanding voltage test, in order to ensure continuity from the MX100 or
equivalent’s earth to the module’s heat sink, install the B8720BB onto the base plate and
fasten with screws.
Measured Item
Criteria
Input terminal to earth terminal
on the MX100, MW100, or equivalent
Current leakage of 2 mA or less at 3700 Vrms AC for 1
minute
Between input terminals
Current leakage of 1 mA or less at 1000 Vrms AC for 1
minute
2.5 Tests of the 10-CH, Medium-Speed Universal Input Module (MX110-UNV-H10)