Yokogawa PC-Based MX100 User Manual
Page 76
2-47
SM MX100-01E
Testing
1
2
3
4
5
6
7
PWM Outupt Settings
Connect to the PC, then enter the following settings using the MX100 standard software.
When using the MX100/DARWIN API for settings, enter settings that are equivalent to
the following.
Item
Setting
Remarks
Enable/disable
Enabled
Action
User output
Span
Need not specify
Output upon power ON
Preset value
Sets the output at power ON as a preset value.
Output upon error
Preset value
Sets the output upon error as a preset value.
Preset value
0.000 %
Sets 0% duty.
Resolution
Depends on the test
Pulse interval
Depends on the test
Other settings
Not required
Using the MW100 user’s manual or communication commands, power cycle the MW100
main unit or equivalent, then measure the output on all channels.
Item
Setting
Remarks
Output type
PWM
Output method
COM
Command output
Output range
Depends on the test
Span
Need not specify
Pulse interval
Depends on the test
Preset value
0.000 %
Sets 0% duty.
Reference channel
Not required
Output Operation Setting
Item
Setting
Power ON
PRESET
Error occurence
PRESET
Criteria (for the MX100 and MW100)
If entering the pulse interval and PWM settings according to the table below, confirm that
the output pulse interval of the specified channels satisfies the criteria.
(Set for each channel. Channels not under test output 0%).
Pulse Interval
Output Setting
Criteria
1 ms
50.000 %
1 ms ±80 ns
5 ms (1 ms ×5)
50.000 %
5 ms ±400 ns
10 ms (10 ms ×1)
50.000 %
10 ms ±800 ns
100 ms (10 ms ×10)
50.000 %
100 ms ±8 s
2.10 8-CH, Medium-Speed PWM Output Module