2 insulation resistance test, 3 withstanding voltage test – Yokogawa PC-Based MX100 User Manual
Page 32
2-3
SM MX100-01E
Testing
1
2
3
4
5
6
7
2.2.1
Protective Earth Continuity Test
Procedure
1.
Wire the instruments as shown below.
Earth continuity
tester
Protective earth
terminal
MX100
(main unit)
Functional
earth terminal
I–
V–
I+
V+
Protective earth terminal
L
N
G
When using an inlet
When using a screw terminal
Center pin
Right-most pin
2.
Measure the resistance between the protective earth terminal and functional earth
terminal. Use 4 wires for this measurement.
Items to Be Checked
That the resistance value does not exceed 0.1 Ω when tested at 60 Hz, 30 A for 10
seconds.
2.2.2
Insulation Resistance Test
Procedure
Perform a test using a DC 500 V insulation resistance tester and confirm that the
following criteria are met.
Terminals Measured
Power Switch
Criteria
AC power terminal to functional earth termina
l*
(Short between L and N)
ON
20 MΩ or higher
Ethernet input terminal to functional earth terminal
(short all pins)
OFF
20 MΩ or higher
* You may use the functional earth terminal as a protective earth terminal.
2.2.3
Withstanding Voltage Test
Procedure
Perform the withstanding voltage test, and confirm that the following criteria are met and
that no abnormalities occur.
Terminals Measured
Power Switch
Criteria
AC power terminal to functional earth terminal
*
(Short between L and N)
ON
1.5 kVAC for 1 minute
Leakage current of 5
mA or less
* You may use the functional earth terminal as a protective earth terminal.
2.2 Tests of the Main Module (MX100)