13. At this point in the auto settling setup
mode, select
OFFST
for auto calibrator
offset.
14. You are asked to correct the auto
calibrator offset. Press
NO
( the default
selection) if the screen message
CUR
-
RENTLY
ENABLED
appears, or press
YES
if the screen reads
CURRENTLY
DISABLED
.
15. Press
ESC
twice to return to the Ana-
lyze mode main menu.
Setup: Clock/Scheduled Span or Scheduled Zero
1. After completing initial start-up, select
SETUP
from the Analyze mode menu to
select the auto settling setup function.
2. Press
SYSTM
to continue.
3. The next screen asks you to enter your
authorization code. Use the
UP
and
DOWN
keys to scroll through the letters/
numbers. Press
SELEC
to enter the
desired value for each space in the code.
Press
ACCPT
when finished.
SLO PE
PERO D
VALVES: ANALYZE C=9:59
T=25C
SELECT AUTO -SETTLE PARAM ETER
TO M O DIFY - PERIO D, SLO PE
OR AUTO CALIBRATOR OFFSET
1
2
3
4
5
O FFST
ESC
YES
VALVES: ANALYZE C=9:59
T=25C
ENABLE AUTO CALIBRATOR OFFSET
CORRECTIO N?
CURRENTLY DISABLED
1
2
3
4
5
NO
ESC
VALVES: ANALYZE
T=25C
C=9:59
FLO W = 150.0 CC/M IN
AUTORANGE: 10 PPM
1
2
3
4
5
RANGE
ALRM R
5.00
PPM 02
SETUP
STDBY
CAL
VALVES: ANALYZE
FLO W = 150.0 CC/M IN
AUTORANGE: 10 PPM
1
2
3
4
5
RANGE
ALRM R
5.00
PPM 02
T=25C
C=9:59
SETUP
STDBY
CAL
VALVES: ANALYZE
T=25C
C=9:59
SELECT SYSTEM PASSW O RD SETUP
- - O R - -
SERIAL M O DEM , ZERO GAS SCRUB
OR NEGATIVE O2 READINGS
1
2
3
4
5
SYSTM
M O D EM
G A S
NegO 2
ESC
VALVES: ANALYZE C = 9:59 T = 25C
ENTER AUTHORIZATION CODE
ENTER YOUR CODE _ _ _