Figure 6-4. Calibration/Span Calibration with External Source

CAL STDBY SETUP RANGE ALRMR

CAL STDBY SETUP RANGE ALRMR

SHOW

ZERO

SPAN

ANALY

GAS

AUTO

DISPLAYS WHEN ANALYZER

WAS LAST SPANNED

OR ZEROED

CHOOSE A

CALIBRATION

METHOD

CONNECT

SPAN

GAS

CHECK

FLOW RATE

NO

ACK

SCREEN DISPLAYS OXYGEN CONTENT

OF SPAN GAS

USE UP AND DOWN KEYS TO SCROLL

TO THE ACTUAL SPAN GAS

CONCENTRATION THEN

ACCPT

SELECT

MANUAL OR

AUTO SETTLING

SCREEN WILL DISPLAY

PRESENT OXYGEN

CONCENTRATION,

FLOW RATE AND

TIME REMAINING.

PRESSING ANY OTHER

KEY BUT ACK WILL

ABORT OUT OF

AUTO SETTLING

CALIBRATION PROCESS

COMPLETED. PRESS

ACK TO RETURN TO

ANALYZE MODE

ACK

YES

IS

SPAN GAS

CONNECTED?

UP

DOWN

ACCPT

ACK

ACK

NOTE: Pressing ESC brings you

back to previous screen.

Span Calibration Using External Gas
Source