After the electrolyte has been purged, it can be added to the sensor using the
following procedure:

1. Remove the plastic screw cap from the anode port on the stainless

steel plate on the top of the sensor body (Figure 3-4.1) and flush out
the sensor by using oxygen-free gas.

Figure 3-4.1. Flushing Out the Ultra Trace Sensor by O

2

-Free Gas

2. Insert the tube attached to the 50 cc syringe into the electrolyte

bottle and slowly fill the syringe with electrolyte.

3. Insert the tube attached to the syringe into the sensor through the

opening on the top stainless steel plate. Slowly pour electrolyte into
the sensor (Figure 3-4.2).

Oxygen-Free
Gas In

Plastic/Teflon

Tube

Gas Out