Figure 3-4.2. Adding Electrolyte to the Ultra Trace Sensor

NOTE: Review the cautions and warnings on the use of electrolyte at

the beginning of this chapter and in the Material Safety Data
Sheets.

Avoid spilling electrolyte on the stainless steel plate. If you spill
any electrolyte on the stainless steel plate, wipe it up with damp
tissue paper. If electrolyte is not completely wiped, a white
residue (potassium carbonate, a reaction product of potassium
hydroxide and carbon dioxide in air) will appear with time.

As you pour, ensure that no large air bubbles remain trapped
inside the sensor. If large bubbles appear, use the syringe to
withdraw electrolyte from the sensor until the bubbles disap-
pear. Return the withdrawn electrolyte to the sensor slowly.

4. Add electrolyte until the level reaches about 1 inch above the yellow

level marker on the front of the Ultra Trace sensor body. DO NOT
add electrolyte above the green level marker.

5. Remove the cadmium anode from the holder and immediately insert

it into the sensor through the opening in the top mounting plate
(Figure 3-5). Ensure that the O-ring is attached to the anode assem-
bly.

TOP

ELECTROLYTE

CELL BODY

ANODE
PORT