Warm Start

Once these parameters are set, they are stored in a battery backed-up, power
failure-safe memory, and the next power-up will automatically use this stored
data (

WARM

START

)

. You can change any of these parameters during operation

except for the authorization code where you will be required to perform a

COLD

START

. Should you lose the code or want to change the code and decide to

COLD

START

the analyzer, previous calibration and alarm set-up data will be

lost, and you will have to span and zero calibrate the analyzer again.

Full Start Versus Quick Start

Every time the analyzer is powered up, the first menu will give you the option
of a Full Start or a Quick Start. Full Start automatically purges the sample,
zero, calibrator and span gas lines. The purge operation is completed in about
two hours. Be sure to use

FULL

START

the very first time you operate the

analyzer.

Q

UICK

START

is used when you are operating the analyzer on a daily basis and

want to disconnect the sample/span gas and connect a new source for continued
normal operation or turned the analyzer OFF/ON switch during troubleshoot-
ing or maintenance.

Q

UICK

START

purges the sample/span line for five minutes

and brings the analyzer quickly on line.

Forcing a COLD START by turning Power OFF/ON

To force a

COLD

START

any time after the very first analyzer power-up, turn

the power OFF, wait for 10 seconds and then hold down the red key on the touch
panel while flipping the analyzer power switch to ON. The very first screen will
indicate

COLD

START

.