Warm Start
Once these parameters are set, they are stored in a battery backed-up, power
failure-safe memory, and the next power-up will automatically use this stored
data (
WARM
START
)
. You can change any of these parameters during operation
except for the authorization code where you will be required to perform a
COLD
START
. Should you lose the code or want to change the code and decide to
COLD
START
the analyzer, previous calibration and alarm set-up data will be
lost, and you will have to span and zero calibrate the analyzer again.
Full Start Versus Quick Start
Every time the analyzer is powered up, the first menu will give you the option
of a Full Start or a Quick Start. Full Start automatically purges the sample,
zero, calibrator and span gas lines. The purge operation is completed in about
two hours. Be sure to use
FULL
START
the very first time you operate the
analyzer.
Q
UICK
START
is used when you are operating the analyzer on a daily basis and
want to disconnect the sample/span gas and connect a new source for continued
normal operation or turned the analyzer OFF/ON switch during troubleshoot-
ing or maintenance.
Q
UICK
START
purges the sample/span line for five minutes
and brings the analyzer quickly on line.
Forcing a COLD START by turning Power OFF/ON
To force a
COLD
START
any time after the very first analyzer power-up, turn
the power OFF, wait for 10 seconds and then hold down the red key on the touch
panel while flipping the analyzer power switch to ON. The very first screen will
indicate
COLD
START
.