Operating the Analyzer

Start-up Checklist

Before you turn the analyzer power on, check the following:

1. Do not begin operation until each of the installation procedures has

been followed and checked (see Chapter 3).

2. Confirm that the analyzer sample gas port is securely connected to

the sample gas source. Check for leaks.

3. The recommended sample gas pressure is between 10 and 30 psig; a

pressure above or below these limits may cause an error in the gas
flow rate which may cause an error in calibration while using the
internal calibrator.

4. If using a span gas to calibrate, the span gas tank should be securely

connected to the span gas port. Span gas pressure should be
between 10 and 30 psig.

SAMPLE
BYPASS
OUT

CALIBRATOR
BYPASS
OUT

WATER

RESERVOIR

RESERVOIR

DRAIN

SPARGER

VENT

SPAN GAS

IN

AIR

IN

SAMPLE IN

SAMPLE/SPAN

VENT

AC POWER IN

5 ALARM CIRCUIT
CONNECTIONS

RS232 SERIAL
PORT (9 PIN D-SUB)
(MALE)

COOLING
FAN

4 SIGNAL CONNECTIONS
(2) 0-1VDC & (2) 4-20MADC

INCLUDING RANGE IDENTIFYING OUTPUT

PARALLEL (PRINTER)
PORT (25 PIN D-SUB)
(FEMALE)

Figure 4-6. Rear Panel Connections