10. After pressing

ACK

, the screen indicates

that the system is calibrating and
displays the message:

PLEASE

WAIT

...

If you try to adjust the span level with
the signal output less than

1

/

3

of the span

set value , you will see the message

"

CELL

TOO

WEAK

"

. If this occurs,

condition the sensor. Pressing

ACK

returns you to the Analyze mode main
menu.

The timer will start to count down.

11. After the calibration process is com-

pleted, the screen displays the present
oxygen content and the flow rate. Press

ACK

to return to the Analyze mode.

12. If you press

ACK

, the system returns to

the Analyze mode, and the screen
displays the oxygen content of the
sample gas and all available menu
options.

VALVES: GAS SPAN

CALIBRATION IN PROGRESS
PLEASE WAIT . . .

1

2

3

4

5

VALVES: SHUTTING DOWN

1

2

3

4

5

TIME REMAINING 00 : 02 : 00

PLEASE WAIT . . .

YES

NO

ESC

VALVES: G AS SPAN

O XYG EN = 5.0 PPM O 2
PRESS ACK TO START THE
ANALYZE M O DE

--- PRESS ACK W HEN READY ---

1

2

3

4

5

ACK

VALVES: ANALYZE

T=25C

C=9:59

FLO W = 150.0 CC/M IN

AUTORANGE: 10 PPM

1

2

3

4

5

RANGE

ALRM R

5.00

PPM 02

SETUP

STDBY

CAL

VALVES: G AS SPAN

% % % % % % % W ARNING % % % % % % %

CELL TO O W EAK

--- PRESS ACK W HEN READY ---