Model 3060e, Teledyne electronic technologies, After pressing – Teledyne 3060e - Ultra Trace oxygen analyzer User Manual
Page 76: Of the span set value , you will see the message, If this occurs, condition the sensor. pressing
![background image](/manuals/345655/76/background.png)
4-28
Model 3060E
Model 3060E
Model 3060E
Model 3060E
Model 3060E
Chapter 4
Chapter 4
Chapter 4
Chapter 4
Chapter 4
Teledyne Electronic Technologies
Analytical Instruments
10. After pressing
ACK
, the screen indicates
that the system is calibrating and
displays the message:
PLEASE
WAIT
...
•
If you try to adjust the span level with
the signal output less than
1
/
3
of the span
set value , you will see the message
"
CELL
TOO
WEAK
"
. If this occurs,
condition the sensor. Pressing
ACK
returns you to the Analyze mode main
menu.
•
The timer will start to count down.
11. After the calibration process is com-
pleted, the screen displays the present
oxygen content and the flow rate. Press
ACK
to return to the Analyze mode.
12. If you press
ACK
, the system returns to
the Analyze mode, and the screen
displays the oxygen content of the
sample gas and all available menu
options.
VALVES: GAS SPAN
CALIBRATION IN PROGRESS
PLEASE WAIT . . .
1
2
3
4
5
VALVES: SHUTTING DOWN
1
2
3
4
5
TIME REMAINING 00 : 02 : 00
PLEASE WAIT . . .
YES
NO
ESC
VALVES: G AS SPAN
O XYG EN = 5.0 PPM O 2
PRESS ACK TO START THE
ANALYZE M O DE
--- PRESS ACK W HEN READY ---
1
2
3
4
5
ACK
VALVES: ANALYZE
T=25C
C=9:59
FLO W = 150.0 CC/M IN
AUTORANGE: 10 PPM
1
2
3
4
5
RANGE
ALRM R
5.00
PPM 02
SETUP
STDBY
CAL
VALVES: G AS SPAN
% % % % % % % W ARNING % % % % % % %
CELL TO O W EAK
--- PRESS ACK W HEN READY ---
1
2
3
4
5
ACK