Ranges:
0-50 parts per billion (PPB) O
2
0-100 PPB O
2
0-1.0 parts per million (PPM) O
2
0-10 PPM O
2
Background Gas:
nitrogen, argon, helium or hydrogen
Accuracy:
±2% of full scale at constant temperature or
±5% of full scale over the operating temperature
range
Response Time:
90% of reading in 75 seconds at constant tempera-
ture
Sensitivity:
0.5 PPB O
2
on 0-50 PPB range
Resolution:
1 PPB O
2
on 0-50 PPB and 0-100 PPB ranges
Operating:
+41 °F to +104 °F (+5 °C to +40 °C)
Temperature Range
Signal Outputs:
For percent of full scale indication:
0-1 VDC non-isolated signal
4-20 mADC isolated signal
For range identification:
0-1 VDC non-isolated signal
4-20 mADC isolated signal
Data Line:
RS-232C serial interface
Alarms:
Five user-programmable absolute reading alarm
setpoints and Form C (SPDT) relay contacts (3A
resistive)
Mounting:
19" relay rack (12.25" high, 18" deep), for general
purpose (nonhazardous) areas
Wetted Surfaces:
316L stainless steel sample system with electropol-
ished wetted surfaces
Connections:
Orbital butt weld and VCR-type fittings
Flow Control:
Provided by a mass flow controller
Recommended Flow: Approximately 150 cc/min
Sample Pressure:
8 to 13 psig, 30 psig maximum
Readouts:
one red 4-digit LED with 1" numeral
one LCD graphics display.
Power Requirements: 100, 120, 220 or 240 VAC
50/60 Hz
Security:
User-programmed authorization entry code
Calibration Gases:
Zero: Integral scrubber makes O
2
-free zero gas
Span: Integral Faradaic calibrator makes a known
span gas. If preferred, a span gas inlet port is
provided to accommodate a certified span gas
*
Specifications subject to change without notice.
Specifications