HIGH CURRENT MODE will automatically activate when:

The oxygen level rises above 10 PPM. This mode will automatically
display 10.00 PPM OVERANGE message.

If the oxygen level rises above 30 PPM for over 2
minutes, a safety valve sequence will commence to
isolate and protect the internal scrubber and Ultra
Trace sensor. The Ultra Trace sensor will reconnect
every 2 minutes to check the oxygen level. If the
oxygen concentration falls below 30 PPM, normal
operation will resume.

If the measured oxygen level remains above 30 PPM
for 60 minutes, the unit will automatically shut down.