NOTE: If you try to use the calibrator with insufficient electrolyte or

without the current source leads connected, the LCD will dis-
play CURRENT FAILURE MODE. If this occurs, refill the
calibrator with calibrator electrolyte, connect the current source
leads, and turn the analyzer OFF and then ON. Proceed with
the Quick Reference: Start Up and perform the calibration
again (see Chapter 4 for details).

Adding Electrolyte to the Calibrator

CAUTION:

Calibration electrolyte is not the same as the Ultra Trace
sensor electrolyte. Calibration electrolyte is a combination of
an aqueous 0.5% CF

3

SO

3

H and an ionic surfactant. Do not

use the aqueous 15% KOH electrolyte (Ultra Trace sensor
electrolyte) in the calibrator. Using the Ultra Trace sensor
electrolyte in the calibrator will seriously damage the calibra-
tor and the sampling system valves.