Full Start Options
Setup for Internal Calibrator Use (Full Start)
6.
If you request the internal calibra-
tor, you are directed to fill the
calibrator with electrolyte. After
the process is completed, press
ACK
to continue or
ESC
to return to
the previous screen.
7.
After pressing
ACK
in step 6, the
next screen gives the option of
connecting the span gas tank.
Press
NO
for internal calibrator
only.
8.
After pressing
NO
in step 7, the
system directs you to connect the
sample gas tank to the appropri-
ate inlet. After the process is
completed, press
ACK
. This en-
sures that the sample gas is
connected to the sample inlet of
the analyzer
9.
After pressing
ACK
in step 8, the
system performs a sample line
purge, then automatically contin-
ues on to step 10.
VALVES: O FF
CAREFULLY FILL THE CALIBRATOR
W ITH A HALF CC OF CALIBRATIO N
ELECTRO LYTE
--- PRESS ACK W HEN READY ---
1
2
3
4
5
ESC
ACK
VALVES: O FF
DO YO U W ANT TO CONNECT THE
OPTIONAL SPAN GAS TANK
AT THIS TIM E?
1
2
3
4
5
NO
ESC
YES
VALVES: O FF
CONNECT A SAM PLE GAS
TO THE SAM PLE PO RT
--- PRESS ACK W HEN READY ---
1
2
3
4
5
ACK
ESC
VALVES: PURG E-TO -SPAN
PURG E IN PRO G RESS
TIM E REM AINING 00 : 01 : 00
PLEASE W AIT . . .
1
2
3
4
5
SKIP