4-16

Model 3060E

Model 3060E

Model 3060E

Model 3060E

Model 3060E

Chapter 4

Chapter 4

Chapter 4

Chapter 4

Chapter 4

Teledyne Electronic Technologies

Analytical Instruments

Setup for Both Span Gas & Internal Calibrator (Full
Start)

15. After saying

YES

to the internal calibra-

tor, you are directed to fill the calibrator
with electrolyte. When done, press

ACK

to continue or

ESC

to return to the

previous screen.

16. After pressing

ACK

in step 15, the next

screen gives the option of connecting
the span gas tank. Press

YES

to use span

gas in addition to the internal calibrator.

17. After pressing

YES

in step 16, the

system directs you to connect the span
gas tank to the appropriate inlet. When
done, press

ACK

to continue.

18. The system directs you to connect the

sample gas tank to the appropriate inlet.
When done, press

ACK

to continue.

19. The system directs you to check the

flow rate for 150 cc/min and to press

ACK

when ready.

NOTE: Mass flow controller is preset to give 150 ± 3 sccm flow. However,

when a pneumatic valve does not open or the gas pressure drops to
a point where the mass flow controller is unable to deliver 150 sccm
flow, the analyzer will automatically shut down and display the
error message FLOW BELOW 20 SCCM. If this occurs, check the
sample/span gas and air pressure. Make sure that they are within
the recommended range.

VALVES: O FF

CAREFULLY FILL THE CALIBRATOR
W ITH A HALF CC OF CALIBRATIO N
ELECTRO LYTE
--- PRESS ACK W HEN READY ---

1

2

3

4

5

ESC

ACK

VALVES: PURG E-FRO M -SPAN

FLO W = 49.5 CC/M IN LO W

1

2

3

4

5

CHECK THE FLO W RATE FO R
150 CC/M IN

--- PRESS ACK W HEN READY ---

ACK

ESC

VALVES: O FF

CONNECT A SAM PLE GAS
TO THE SAM PLE PO RT

--- PRESS ACK W HEN READY ---

1

2

3

4

5

ACK

ESC

VALVES: O FF

CONNECT THE SPAN GAS TANK
TO THE SPAN INLET

1

2

3

4

5

--- PRESS ACK W HEN READY ---

ACK

ESC

VALVES: OFF

DO YOU WANT TO CONNECT THE
OPTIONAL SPAN GAS TANK
AT THIS TIME?

1

2

3

4

5

NO

ESC

YES