Use COLD START when:

The Instrument is turned on for the first time.

A new authorization code is desired.

You want to reset all parameters to the default factory set values.

Use WARM START when:

The Instrument is turned on after a lengthy period of non-use and
you want to retain previously set parameters.

After filling the Ultra Trace sensor with electrolyte.

You want to purge the gas lines.

FULL START:

A 2 hour purge of all gas lines

(sample,zero, span and calibrator).

WARM START: A 2 minute purge of sample
line only. Used when connecting a new sample
source or when instrument has been turned off
for a brief period.

Use ANALYZE MODE when:

You want to measure the oxygen concentration in a sample.

You want to calibrate the analyzer.

You want to setup the instrument for automatic calibration cycles.

CALIBRATION: Set the zero and span of the
instrument using internal or external sources.

SETUP:

Schedule intervals for AUTO ZERO and AUTO
SPAN calibration cycles.

Enter values for AUTO SETTLING.

Enter values for signal FILTERing.

Set the display for positive O

2

.

Set the ALARM parameters.

Set the MODEM configturation.

RANGE: Switch between MANUAL range and
AUTO RANGE.

STANDBY:

Maintain operational status for instrument when
no gas is flowing .

You want to disconnect a gas line for mainte-
nance or reconnect another source.