Operations – Teledyne 3060e - Ultra Trace oxygen analyzer User Manual
Page 73
Teledyne Electronic Technologies
Analytical Instruments
4-25
Operations
Operations
Operations
Operations
Operations
Ultra
Ultra
Ultra
Ultra
Ultra T
T
T
T
Trace Oxyg
race Oxyg
race Oxyg
race Oxyg
race Oxygen Anal
en Anal
en Anal
en Anal
en Analyz
yz
yz
yz
yzer
er
er
er
er
Span Calibration Using an External Span Gas
1. Once you are in the Analyze mode, the
screen displays all the options available
in this mode. Press
CAL
to calibrate.
2. The screen directs you to select from
the Calibration mode menu displayed.
Press
SPAN
to adjust the span level.
Pressing
SHOW
displays when the
analyzer was last spanned or zeroed,
and the results.
3. When you select
SPAN
, the screen tells
you to choose a calibration method.
Press
GAS
to calibrate with an external
span gas.
4. After pressing
GAS
, the screen asks
you if a span gas is connected to the
span port.
5. If the span gas is connected, press
YES
.
The screen displays the flow rate and
asks you to check it for 150 cc/min.
Press
ACK
when ready.
If the span gas is not connected, the
display will ask you to connect it, check
the flow rate, and press
ACK
when
ready.
VALVES: ANALYZE
T=25C
C=9:59
FLO W = 150.0 CC/M IN
AUTORANGE: 10 PPM
1
2
3
4
5
RANGE
ALRM R
5.00
PPM 02
SETUP
STDBY
CAL
VALVES: ANALYZE
SELECT CALIBRATIO N FUNCTIO N
1
2
3
4
5
ANALY
SPAN
ZERO
SHO W
VALVES: ANALYZE
SELECT SPAN CALIBRATIO N
USING EITHER THE INTERNAL
CALIBRATOR OR AN EXTERNAL
SPAN GAS
1
2
3
4
5
ESC
G A S
INTRN
VALVES: ANALYZE
IS A SPAN GAS CONNECTED
TO THE SPAN PO RT?
1
2
3
4
5
NO
ESC
YES
VALVES: GAS SPAN
OXYGEN = 5.00 PPM O2
FLOW = 150 CC/MIN
1
2
3
4
5
CHECK THE FLOW RATE FOR
150 CC/MIN
--- PRESS ACK WHEN READY ---
ACK
ESC