Chapter 5

Chapter 5

Chapter 5

Chapter 5

Chapter 5

Teledyne Electronic Technologies

Analytical Instruments

YES

NO

FAILURE

CORRECTED

FLOW FAILURE

<20cc/min

150cc/MIN OPTIMUM

PROCEED

CALL

SERVICE

YES

NO

FAILURE

CORRECTED

PRESS THE START KEY

TO RESET THE ANALYZER

TO THE FACTORY

SETTINGS

CHECK AIR SUPPLY
PRESSURE
(75-80 PSIG)

CHECK SAMPLE SPAN
GAS PRESSURE
(10-30 PSIG)

1

2

CHECK MFC
CABLE &
CONNECTIONS

CHECK BACKGROUND
GAS SELECTION KNOB
ON FRONT PANEL IS SET
TO APPROPRIATE GAS

NO FLOW INDICATED

1

2

PRESS THE START KEY

TO RESET THE ANALYZER

TO THE FACTORY

SETTINGS

Table 5-6. Flow Failure Less than 20cc