Installation Checklist

After installing the system, confirm the following:

The sample gas connection to the analyzer sample inlet port is not
leaking.

The sample gas pressure must be between 10 and 30 psig.

The span gas pressure must be between 10 and 30 psig.

There are no leaks between the span gas tank and the analyzer span
gas inlet port when a span gas is used to calibrate.

The air supply connection to the analyzer port is not leaking.

The air supply pressure must be between 75 and 80 psig.

The vent tubes must be properly installed and should be at least

1

/

4

"

outer diameter. (Tubing with less than

1

/

4

" O.D. causes back-

pressure that forces gas to flow through the sensing membrane of
the Ultra Trace sensor, generating signal noise.)

The electrolyte level in the Ultra Trace sensor must be between the
green and yellow marker levels.

CAUTION:

Protective equipment including but not limited to gloves,
safety glasses, face shield and rubber apron must be worn
while handling electrolyte. The safety of the people in the
vicinity of the area in which the electrolyte is being
handled must be given similar consideration. Please refer
to Material Safety Data Sheets to learn about potential
hazards and corrective action in case of accident.