Figure 6-6. Calibration/Zero Calibration

CAL STDBY SETUP RANGE ALRMR

CAL

ZERO

THE SCREEN DISPLAYS WHETHER

THE ANALYZER IS USING THE

INTERNAL SCRUBBER

OR AN EXTERNAL ZERO GAS

PRESS ACK WHEN THE BACKGROUND

SIGNAL STABILIZES AND IS WITHIN

THE APPLICABLE LIMIT

SCREEN WILL DISPLAY PRESENT
OXYGEN CONCENTRATION, FLOW RATE
AND TIME REMAINING

MAN

CALIBRATION PROCESS COMPLETED

PRESS ACK TO RETURN TO ANALYZE MODE

PRESSING ANY KEY WILL ABORT
AUTO SETTLING

PRESSING ACK AFTER
ABORTING AUTO SETTLING
RETURNS YOU TO ANALYZE MODE

If the background level does not
fall below the acceptable limit after
5-7 days, replace the electrolyte with
prepurged electrolyte. Refer to Sections
3.3.1 and 4.3.7.2 for more details

NOTE:

AFTER AUTO SETTLING, THE SYSTEM
WILL AUTOMATICALLY RETURN
TO THE ANALYZE MODE

NOTE: Pressing ESC brings you

back to previous screen.

Zero Calibration