High Current Output Mode

1. If

O

2

> 10

PPM

If the oxygen (O

2

) concentration level

exceeds 10 PPM, the system will enter
High Current mode. The screen will
show

HI

-

CURRENT

as the range mode,

and

10.00

PPM

O

2

OVERRANGE

as

the concentration.

2. If

O

2

> 30

PPM

If the O

2

concentration level exceeds 30

PPM for longer than two minutes, the
scrubber valves PV6 and PV7 will close
to protect the scrubber, and the Ultra
Trace sensor will be disconnected.
Every 10 minutes thereafter, the Ultra
Trace sensor will be connected for 2
minutes while O

2

measurements are

taken.

If the O

2

level returns to levels below 30

PPM, the scrubber valves PV6 and PV7
will open, and the Ultra Trace sensor
will continue to monitor O

2

.

If the O

2

level remains above 30 PPM

for 60 minutes, the system will shut
down automatically. This is to prevent
damage to the Ultra Trace sensor and
scrubber. Check for leaks around gas
connections and try to restart the
analyzer.

VALVES: ANALYZE

T=25C

C=9:59

FLO W = 49.5 CC/M IN

AUTORANGE: HI-CURRENT

1

2

3

4

5

RANGE

ALRM R

10.00

OVERRANGE
PPM 02

SETUP

STDBY

CAL

10.00 blinks in High Current m ode