beautypg.com

Model 3060e, Ultra trace sensor electrical connections – Teledyne 3060e - Ultra Trace oxygen analyzer User Manual

Page 34

background image

3-8

Model 3060E

Model 3060E

Model 3060E

Model 3060E

Model 3060E

Chapter 3

Chapter 3

Chapter 3

Chapter 3

Chapter 3

Teledyne Electronic Technologies

Analytical Instruments

Figure 3-5. Inserting the Anode

Ultra Trace Sensor Electrical Connections

There are two wires to be connected so that the sensor can send signals to the
central processor to generate the O

2

content data, and two wires to send

temperature data to the processor.

The cathode terminal is located at the bottom of the sensor body and connected
to the red wire. The anode connection is on top of the sensor and is connected
to the black wire. The thermistor terminal is mounted on the stainless steel plate
on the top of the sensor body. With the exception of the anode, each of these
terminals is already connected to the signal processing unit when the analyzer
is shipped.

The positive and negative current source leads for the calibrator are located on
the top of the calibrator assembly (Figure 3-6).

After installing the analyzer, examine all of the connectors to ensure that they
are firmly secured to their respective terminals.

TOP

ANODE
PORT

ELECTROLYTE

CELL BODY