Span Calibration Using the Internal Calibrator

1. From the Analyze mode, press

CAL

to

span calibrate using internal calibrator.

2. The screen directs you to select from

the Calibration mode menu displayed.
Press

SPAN

to calibrate.

3. When you select

SPAN

, the screen tells

you to choose a calibration method.
Press

INTERN

for the Internal Calibra-

tion mode.

If you have warm started, go to step 6.

If the calibrator has been filled previous-
ly, the screen asks: “

DO

YOU

WISH

TO

FILL

THE

CALIBRATOR

AT

THIS

TIME

4. If you answer

YES

, the screen instructs

you to fill the calibrator with the correct
amount of calibration electrolyte (in-
structions below). Press

ACK

to

continue or

ESC

to return to the previ-

ous menu.

Add

1

/

2

cc of calibration electrolyte. One half cc of calibration electrolyte should

be sufficient to use the calibrator for up to 10 hours, either continuously or off-
and-on. The calibrator display (in hh:mm) on the screen keeps track of the time
remaining.

If you try to use the internal calibrator without sufficient electrolyte, or without
the current source cable connected, the analyzer will display the message
CURRENT FAILURE. If this occurs, press ACK to return to the Analyze
mode. Fill the calibrator and start the calibration again. For more detail, see
Conditioning the Ultra Trace Sensor earlier in this chapter.

VALVES: ANALYZE
SELECT SPAN CALIBRATIO N
USING EITHER THE INTERNAL
CALIBRATOR OR AN EXTERNAL
SPAN GAS

1

2

3

4

5

ESC

G AS

INTRN

VALVES: ANALYZE

T=25C

C=9:59

FLO W = 150.0 CC/M IN

AUTORANGE: 10 PPM

1

2

3

4

5

RANGE

ALRM R

5.00

PPM 02

SETUP

STDBY

CAL

VALVES: CALIBRATOR ANALYZE

--- PRESS ACK W HEN READY ---

CAREFULLY FILL THE CALIBRATOR
W ITH A HALF CC OF CALIBRATIO N
ELECTRO LYTE

1

2

3

4

5

ESC

ACK