31. The Ultra Trace sensor is reconnected.
After the sensor connects, the system
automatically returns you to the Ana-
lyze mode main menu.
•
From the Analyze mode you can
calibrate, set up the alarms, modem,
etc., set the range, or go to stand-by.
Quick Start Mode
1. Turn the power switch
ON
. The first
screen displays the name of the com-
pany, instrument model numbers and
software version. You are asked to wait
for start-up initialization.
2. After a few minutes, the screen displays
the options
FULL
or
QUICK
START
mode. Press
QUICK
for
QUICK
START
.
3. After pressing
QUICK
the screen directs
you to select the background gas using
the knob on the front panel. Press
ACCPT
after selection. Press
ESC
to
return to the previous screen.
4. After you press
ACCPT
to select a
background gas, the screen directs you
to connect the sample gas to the sample
port. Press
ACK
or
ESC
to continue.
VALVES: ANALYZE
T=25C
FLO W = 150.0 CC/M IN
AUTORANGE: 10 PPM
1
2
3
4
5
RANGE
ALRM R
5.00
PPM 02
SETUP
STDBY
CAL
VALVES: ANALYZE
CONNECTING SENSOR
PLEASE WAIT . . .
1
2
3
4
5
VALVES: OFF
1
2
3
4
5
WARNING: IF THIS IS THE FIRST
SETUP THE FULL STARTUP
IS REQUIRED
FULL
QUICK
VALVES: OFF
1
2
3
4
5
CONNECT A SAMPLE GAS
TO THE SAMPLE PORT
. . . PRESS ACK WHEN READY . . .
ACK
ESC