You may abort auto settling by pressing
any key. Pressing

ACK

after abort will

return you to the Analyze mode main
menu.

5. In the Zero Calibration mode, the

screen displays whether the analyzer is
using the internal scrubber to automati-
cally create the zero gas or if an external
zero gas has been chosen. (See Setup
Mode: Zero Gas.) Press

ACK

when the

background signal is stable and within
the applicable limit or

ESC

to return to

the Analyze mode.

NOTE: The microprocessor will offset the raw background signal equiva-

lent to -10 to 150 PPB. If the background signal is outside the
acceptable limit, wait until the output reaches to within the
acceptable limit.

6. When background reaches within the

aceptable value, go to step 9. If

ACK

is

pressed while the background signal is
greater than the maximum allowable
value, the screen indicates

CANNOT

ZERO

.

BACKGROUND

LEVEL

MORE

THAN

THE

ACCEPTABLE

VALUE

.

NOTE: Higher background is primarily due to the presence of dissolved

oxygen in the solution. The time required for the removal of
dissolved oxygen from the electrolyte will depend on the initial
concentration of oxygen in the electrolyte. With a well prepurged
electrolyte, the output in the Zero mode will drop below the
acceptable limit within 2–4 days. If the background signal does not
fall below the aceptable limit after 5–7 days, replace the electrolyte
with freshly prepurged electrolyte. Replacing the electrolyte re-
moves the dissolved oxygen from the Ultra Trace sensor and the
background will will drop below the acceptable limit quickly.

VALVES: ZERO , INT SCRUB

% % % % W ARNING % % % %

USER ABO RTED AUTO SETTLE M O DE

--- PRESS ACK W HEN READY ---

1

2

3

4

5

ACK

ESC

VALVES: ZERO , INT SCRUB
O XYG EN = 5.0 PPB O 2
FLO W = 150.0 CC/M IN
W ARNING : THE BACKG RO UND LEVEL
M UST BE > - 10 PPB AND < 150 PPB
TO ZERO . PRESS ACK W HEN STABLE

1

2

3

4

5

ACK

ESC

VALVES: ZERO ING

T=25C

O XYG EN = 0.5 PPM O 2
FLO W = 150 CC/M IN
CANNOT ZERO. BACKGROUND
LEVEL M O RE THAN 150 PPB
ACK TO CHANGE ELECTROLYTE

1

2

3

4

5

ACK

ESC