Teledyne 3060e - Ultra Trace oxygen analyzer User Manual

Page 30

background image

3-4

Model 3060E

Model 3060E

Model 3060E

Model 3060E

Model 3060E

Chapter 3

Chapter 3

Chapter 3

Chapter 3

Chapter 3

Teledyne Electronic Technologies

Analytical Instruments

2. Slip the Teflon tubing (mounted inside the sensor) onto the

1

/

8

"

stainless steel tube at the bottom of the stainless steel top plate (see
Figure 3-2).

Figure 3-2. Tube Alignment

3. Push the sensor all the way up against the top stainless steel plate.

4. Finger-tighten the side and then the bottom bolts with your hand.

5. Using the rachet (RL547) supplied, tighten each of the four bolts to

secure the sensor to the top plate.

6. Using the rachet (RL547) supplied, tighten each of the four bolts on

the side plate.

NOTE: Make sure that the O-rings are in their grooves on the top and side

of the sensor before tightening the bolts.

Adding Electrolyte to the Ultra Trace Sensor

CAUTION:

Protective equipment including but not limited to gloves,
safety glasses, face shield and rubber apron must be worn
while handling electrolyte. The safety of the people in the
vicinity of the area in which the electrolyte is being handled
must be given similar consideration. Please refer to Material
Safety Data Sheets to learn about potential hazards and
corrective action in case of accident.