4. When adding electrolyte to the calibrator, do not overfill.

5. When hooking up electrical connections, do not connect the output

to the chassis.

6. Do not begin operating the analyzer until you have completed all of

the installation procedures.

Ultra Trace Sensor Installation

The analyzer is usually shipped with the sensor installed. If you need to install
or replace the sensor assembly before using the analyzer, use the procedure
below.

1. Prepare the analyzer casing for the sensor. Place four of the short

bolts in the holes on the side of the sensor and four of the long bolts
in the holes at the bottom of the sensor and bring the sensor between
the top and side stainless steel plates (see Figure 3-1). Leave all
eight bolts loose until Step 3.

Figure 3-1. Ultra Trace Sensor Bolt Installation

SIDE BO LTS

BASE BO LTS

O N

POWER

CONTRO

L MODU

LE

PPB

OXYG

EN

PPM

O XYGEN

TELEDYN

E

ANALYTIC

AL INSTR

UMENTS