Model 3060e, Setup: h, O reservoir – Teledyne 3060e - Ultra Trace oxygen analyzer User Manual

Page 100: Teledyne electronic technologies, Keys to scroll through the letters/ numbers. press, When finished. 4. press

background image

4-52

Model 3060E

Model 3060E

Model 3060E

Model 3060E

Model 3060E

Chapter 4

Chapter 4

Chapter 4

Chapter 4

Chapter 4

Teledyne Electronic Technologies

Analytical Instruments

Setup: H

2

O Reservoir

NOTE: Some Ultra Trace Oxygen Analyzers were built with a toggle switch

mounted in the sensor compartment as shown in Figure 3-6. The
purpose of this switch is to toggle off and on the solenoid valve that
controls the replenishing of water to the cell and dislodge any
trapped gas that may be residing in this line. This section describes
the software version of the same valve switching operation. If your
instrument has the toggle switch installed, your software does not
contain this feature. Disregard this section and see Chapter 3,
paragraph, Level Detector and Solenoid Switch for details regard-
ing the manual switching of the solenoid valve.

1. After completing initial start-up, press

SETUP

from the Analyze mode menu to

select system setup.

2

Press

SYSTM

to continue.

3. The next screen asks you to enter your

authorization code. Use the

UP

and

DOWN

keys to scroll through the letters/

numbers. Press

SELEC

to enter the

desired value for each space in the code.
Press

ACCPT

when finished.

4. Press

MORE

to view additional system

setup options, or

ESC

to return to the

main menu.

VALVES: ANALYZE

C=9:59 T=25C

SELECT SYSTEM PASSW O RD SETUP

- - O R - -

SERIAL M O DEM , ZERO GAS SCRUB

OR NEGATIVE O2 READINGS

1

2

3

4

5

SYSTM

M O DEM

G AS

NegO 2

ESC

VALVES: ANALYZE C = 9:59 T = 25C

ENTER AUTHORIZATION CODE

ENTER YOUR CODE _ _ _

VALVES: ANALYZE

C=9:59 T=25C

- - - - SYSTEM SETUP - - - -

CLOCK: SET TIM E AND DATE

ALRM S: SET LIM ITS, TYPES

SETTL: SETUP AUTO SETTLING

1

2

3

4

5

ALRM S

CLOCK

ESC

SETTL

M O RE