Span calibration using internal calibrator, Model 3060e, Teledyne electronic technologies – Teledyne 3060e - Ultra Trace oxygen analyzer User Manual

Page 132

background image

6-8

Model 3060E

Model 3060E

Model 3060E

Model 3060E

Model 3060E

Chapter 6

Chapter 6

Chapter 6

Chapter 6

Chapter 6

Teledyne Electronic Technologies

Analytical Instruments

Figure 6-5. Calibration/Span Calibration Using Internal Calibrator

CAL STDBY SETUP RANGE ALRMR

CAL

SPAN

INTRN

YES

ACK

NO

CHOOSE A

CALIBRATION

METHOD

DO YOU

WISH TO
FILL THE

CALIBRATOR

FILL

CALIBRATOR

SET

FLOW RATE

CALIBRATOR PURGE (5 MIN)

SCREEN DISPLAYS PRESENT
OXYGEN CONTENT AND
FLOW RATE

SCREEN DISPLAYS OXYGEN CONTENT
OF SPAN GAS
USE UP AND DOWN KEYS TO SCROLL TO
THE ACTUAL SPAN GAS CONCENTRATION
THEN PRESS ACCPT

MAN

THE SCREEN DISPLAYS THE

OXYGEN CONCENTRATION

PRESS ACK WHEN THE OXYGEN

LEVEL STABILIZES

(UP TO 30 MIN)

SCREEN WILL DISPLAY PRESENT
OXYGEN CONCENTRATION, FLOW RATE
AND TIME REMAINING

PRESSING ANY OTHER KEY BUT ACK
WILL ABORT OUT OF AUTOSETTLING

CALIBRATION PROCESS COMPLETED
PRESS ACK TO RETURN TO ANALYZE MODE

CALIBRATION PROCESS COMPLETED

PRESS ACK TO RETURN TO ANALYZE MODE

NOTE: Pressing ESC brings you

back to previous screen.

Span Calibration Using Internal Calibrator