beautypg.com

Model 3060e – Teledyne 3060e - Ultra Trace oxygen analyzer User Manual

Page 82

background image

4-34

Model 3060E

Model 3060E

Model 3060E

Model 3060E

Model 3060E

Chapter 4

Chapter 4

Chapter 4

Chapter 4

Chapter 4

Teledyne Electronic Technologies

Analytical Instruments

4. If you have chosen Auto settling, the

screen displays the present oxygen
concentration and the flow rate. It also
checks whether the background signal is
within the acceptable range.

Before the processor begins to check
for the slope of the baseline, a delay of 5
minutes occurs and is used to counter
any spikes that might occur during
mode switching. After an initial delay,
the screen displays the time left in
seconds while the system waits for
settling. After auto settling, the system
will automatically return you to the
Analyze mode main menu.

OXYGEN CONCENTRATION

TIME

Base Line

Press ACK here to adjust span

Figure 4-7. Adjusting Span Sensitivity When Spiking Occurs on the Signal

VALVES: ZERO , INT SCRUB
O XYG EN= 5.0 PPM O 2
FLO W = 150.0 CC/M IN
SETTLING DELAY = 299 SECS
W AIT FO R O 2 SETTLING
OR ANY KEY TO ABORT

1

2

3

4

5