Model 3060e, Teledyne electronic technologies, When you press – Teledyne 3060e - Ultra Trace oxygen analyzer User Manual

Page 70

background image

4-22

Model 3060E

Model 3060E

Model 3060E

Model 3060E

Model 3060E

Chapter 4

Chapter 4

Chapter 4

Chapter 4

Chapter 4

Teledyne Electronic Technologies

Analytical Instruments

9. When you press

ACK

, the screen indi-

cates that the sample line is being
purged and displays the time remaining
until the process is complete.

10. After 5 minutes, the Ultra Trace sensor

is connected and the screen indicates
that the system will wait a maximum of
5 minutes for an oxygen level of 30
PPM or less.

11. If the oxygen content is greater than 30

PPM, the system automatically goes
into 30-minute Auto-shutdown mode. If
the oxygen content is still greater than
30 PPM after 30 minutes, the system
shuts down.

NOTE: If the system shuts down due to high oxygen concentration, check

for leaks around gas connections. Check that gas concentrations
are not more than 30 PPM. Check that there is no back pressure
in the analyzer vent lines. Repeat Full Start.

When the oxygen content is less than 30
PPM, the screen displays the oxygen
content reading and directs you to open
MV1 to begin sparging the electrolyte.
Press

ACK

or

ESC

to continue.

12. When you press

ACK

, the screen displays

the oxygen content and all the mode
options available after Quick Start.

VALVES: FLOW SET

PURGE IN PROGRESS

1

2

3

4

5

TIME REMAINING 00 : 05 : 00

PLEASE WAIT . . .

SKIP

VALVES: FLOW SET

WAITING FOR 30 PPM OR LESS

1

2

3

4

5

OXYGEN =
TIME REMAINING 00 : 05 : 00

PLEASE WAIT . . .

SKIP

VALVES: FLOW SET
WARNING: CANNOT ACHIEVE 30 PPM
OR LESS. CHECK VALVES FOR
LEAKS. AUTO SHUTDOWN STARTED
OXYGEN = 10.00 PPM O2

1

2

3

4

5

TIME REMAINING 00 : 30 : 00

SKIP

VALVES: SHUTTING DOWN

1

2

3

4

5

TIME REMAINING 00 : 02 : 00

PLEASE WAIT . . .

YES

NO

ESC

VALVES: ANALYZE

--- PRESS ACK W HEN READY ---

O XYG EN = 5.0 PPM O 2
FLO W = 150.0 CC/M IN
G ENTLY O PEN M V1 1 DIVISIO N

1

2

3

4

5

ACK

ESC

VALVES: ANALYZE

T=25C

FLO W = 150.0 CC/M IN

AUTORANGE: 10 PPM

1

2

3

4

5

RANGE

ALRM R

5.00

PPM 02

SETUP

STDBY

CAL