Chapter 4
Chapter 4
Chapter 4
Chapter 4
Chapter 4
Teledyne Electronic Technologies
Analytical Instruments
9. When you press
ACK
, the screen indi-
cates that the sample line is being
purged and displays the time remaining
until the process is complete.
10. After 5 minutes, the Ultra Trace sensor
is connected and the screen indicates
that the system will wait a maximum of
5 minutes for an oxygen level of 30
PPM or less.
11. If the oxygen content is greater than 30
PPM, the system automatically goes
into 30-minute Auto-shutdown mode. If
the oxygen content is still greater than
30 PPM after 30 minutes, the system
shuts down.
NOTE: If the system shuts down due to high oxygen concentration, check
for leaks around gas connections. Check that gas concentrations
are not more than 30 PPM. Check that there is no back pressure
in the analyzer vent lines. Repeat Full Start.
•
When the oxygen content is less than 30
PPM, the screen displays the oxygen
content reading and directs you to open
MV1 to begin sparging the electrolyte.
Press
ACK
or
ESC
to continue.
12. When you press
ACK
, the screen displays
the oxygen content and all the mode
options available after Quick Start.
VALVES: FLOW SET
PURGE IN PROGRESS
1
2
3
4
5
TIME REMAINING 00 : 05 : 00
PLEASE WAIT . . .
SKIP
VALVES: FLOW SET
WAITING FOR 30 PPM OR LESS
1
2
3
4
5
OXYGEN =
TIME REMAINING 00 : 05 : 00
PLEASE WAIT . . .
SKIP
VALVES: FLOW SET
WARNING: CANNOT ACHIEVE 30 PPM
OR LESS. CHECK VALVES FOR
LEAKS. AUTO SHUTDOWN STARTED
OXYGEN = 10.00 PPM O2
1
2
3
4
5
TIME REMAINING 00 : 30 : 00
SKIP
VALVES: SHUTTING DOWN
1
2
3
4
5
TIME REMAINING 00 : 02 : 00
PLEASE WAIT . . .
YES
NO
ESC
VALVES: ANALYZE
--- PRESS ACK W HEN READY ---
O XYG EN = 5.0 PPM O 2
FLO W = 150.0 CC/M IN
G ENTLY O PEN M V1 1 DIVISIO N
1
2
3
4
5
ACK
ESC
VALVES: ANALYZE
T=25C
FLO W = 150.0 CC/M IN
AUTORANGE: 10 PPM
1
2
3
4
5
RANGE
ALRM R
5.00
PPM 02
SETUP
STDBY
CAL