2. Press

SYSTM

to continue auto settling

setup.

3. The next screen asks you to enter your

authorization code. Use the

UP

and

DOWN

keys to scroll through the

letters/numbers. Press

SELEC

to enter

the desired value for each space in the
code. Press

ACCPT

when finished.

4. Press

SETTL

to go on to set up the auto

settling, or

ESC

to return to the main

menu.

5. You are given the option of modifying

the auto settling parameters. Select

PEROD

to modify the period for set-

tling.

6. The next screen informs you that the

system is set up for low filter settling of
1–15 minutes and medium or high filter
settling of 6 minutes. Press

ACCPT

to

accept the default settling period (5
minutes), or scroll through the time
values by pressing

UP

or

DOWN

.

7. To continue modifying the auto settling

parameters, select

SLOPE

to modify the

slope for settling.

VALVES: ANALYZE

T=25C

C=9:59

SELECT SYSTEM PASSW O RD SETUP

- - O R - -

SERIAL M O DEM , ZERO GAS SCRUB

OR NEGATIVE O2 READINGS

1

2

3

4

5

SYSTM

M O D EM

G A S

NegO 2

ESC

VALVES: ANALYZE C = 9:59 T = 25C

ENTER AUTHORIZATION CODE

ENTER YOUR CODE _ _ _

VALVES: ANALYZE

C=9:59 T=25C

- - - - SYSTEM SETUP - - - -

CLOCK: SET TIM E AND DATE

ALRM S: SET LIM ITS, TYPES

SETTL: SETUP AUTO SETTLING

1

2

3

4

5

ALRM S

CLOCK

ESC

SETTL

M O RE

SLO PE

PERO D

VALVES: ANALYZE C=9:59

T=25C

SELECT AUTO -SETTLE PARAM ETER
TO M O DIFY - PERIO D, SLO PE
OR AUTO CALIBRATOR OFFSET

1

2

3

4

5

O FFST

ESC

VALVES: ANALYZE C=9:59

T=25C

SET THE PERIO D FO R SETTLING
LO W FILTER

1-15 M INUTES

PERIO D = 5 M INUTES

M ED, HI FILTER = 6 M INUTES

1

2

3

4

5

ESC

UP

DOW N

ACCPT

SLO PE

PERO D

VALVES: ANALYZE C=9:59

T=25C

SELECT AUTO -SETTLE PARAM ETER
TO M O DIFY - PERIO D, SLO PE
OR AUTO CALIBRATOR OFFSET

1

2

3

4

5

O FFST

ESC