Setup: Zero/Span Gas Source
The Analyzer can be zeroed with zero gas generated by the built-in zero gas
scrubber, or by using an external zero gas. Similarly, the Analyzer can be
spanned by using the internal Faradaic calibrator or an external certified span
gas.
1. The factory default for Zero and Span
gas source is internal. In order to
choose external Zero or Span gas press
SETUP
from the Analyze mode.
2. Press
GAS
to continue.
3. You can now select between configur-
ing zero or span gas source.
4. If you press
ZERO
, the display gives you
the option of choosing either the inter-
nal scrubber or an external scrubber.
The last line indicates the present
selection. Then,
INT
or
EXT
returns you
to the screen in step 2.
5. If you press
SPAN
, you can choose
between the internal calibrator and an
external span gas. If you press
INT
, the
screen prompts you to fill the calibrator
with electrolyte, and
ACK
returns you to
the screen in step 2. If you press
EXT
,
the system returns you to the screen in
step 2.
VALVES: ANALYZE
T=25C
C=9:59
FLO W = 150.0 CC/M IN
AUTORANGE: 10 PPM
1
2
3
4
5
RANGE
ALRM R
5.00
PPM 02
SETUP
STDBY
CAL
VALVES: ANALYZE
T=25C
C=9:59
SELECT SYSTEM PASSW O RD SETUP
- - O R - -
SERIAL M O DEM , ZERO GAS SCRUB
OR NEGATIVE O2 READINGS
1
2
3
4
5
SYSTM
M O DEM
G AS
NegO 2
ESC
VALVES: ANALYZE
T=25C
C=9:59
INTERNAL/EXTERNAL SELECTION
FO R ZERO GAS AND SPAN GAS
1
2
3
4
5
SPAN
ZERO
ESC
VALVES: ANALYZE
T=25C
C=9:59
SELECT INTERNAL SCRUBBER OR
EXTERNAL SCRUBBER FO R ZERO GAS
INTERNAL SCRUBBER NOW IN USE
1
2
3
4
5
INT
EXT
ESC
VALVES: ANALYZE
T=25C
C=9:59
SELECT INTERNAL CALIBRATOR OR
EXTERNAL SPAN GAS FO R SPAN CAL
INTERNAL CALIBRATOR NOW IN USE
1
2
3
4
5
INT
EXT
ESC