Teledyne 3060e - Ultra Trace oxygen analyzer User Manual
Page 31
Teledyne Electronic Technologies
Analytical Instruments
3-5
Installation
Installation
Installation
Installation
Installation
Ultra
Ultra
Ultra
Ultra
Ultra T
T
T
T
Trace Oxyg
race Oxyg
race Oxyg
race Oxyg
race Oxygen Anal
en Anal
en Anal
en Anal
en Analyz
yz
yz
yz
yzer
er
er
er
er
Prior to adding electrolyte to the Ultra Trace sensor, the electrolyte must be
purged with nitrogen (N
2
) to remove most of the dissolved oxygen (O
2
). This
will reduce the time required to remove the dissolved O
2
from the electrolyte
by sparging it with N
2
inside the sensor and hence, shorten the start-up time. To
purge the electrolyte:
1. Prepare the sensor electrolyte by following procedure MP-A56133
included in the Accessory Kit.
2. Open the cap of the electrolyte bottle.
3. Insert the fritted glass tube (P/N T1052) through the opening in the
bottle and bubble O
2
-free N
2
through the electrolyte. The flow rate
of the N
2
should be adjusted so that a large number of small bubbles
can be seen dispersing through the electrolyte.
Figure 3-3. Electrolyte Bottle with Fritted Tube
4. Purge the electrolyte for about one hour.
NOTE: TET/AI recommends that the sensor be purged before adding
electrolyte. This is done by lowering a plastic/Teflon tube (clean
and free of any oil/grease and/or particulates) into the sensor
through the opening on the top of the stainless steel mounting plate
and flushing the sensor with O
2
-free gas. Add electrolyte into the
sensor while flushing the sensor with O
2
-free gas. This will
eliminate the O
2
from the sensor before the addition of prepurged
electrolyte. This procedure will substantially reduce the downtime
associated with clean-up at the first startup of the analyzer.
Nitrogen
Teflon/Nylon
tube
Fritted tube
P/N T1052
Electrolyte
bottle
P/N BTL-P-32O Z
KO H
electrolyte
Nitrogen
bubbles