10. The system directs you to check

the flow rate for 150 cc/min and
to press

ACK

when ready.

With recommended sample
gas pressure (10–30 psig), the
mass flow controller will
automatically control the flow
at 150 ± 4 sccm. You may
check the flow by connecting a
bubblometer at the analyzer
sample vent line to insure that
there is proper flow.