Setup for Span Gas Use (Full Start)

11. After pressing

NO

in step 5, the system

directs you to connect the span gas tank
to the appropriate inlet. After complet-
ing the process, press

ACK

to continue.

12. The system directs you to connect the

sample gas tank to the appropriate inlet.
When done, press

ACK

to continue.

13. The system directs you to check the

flow rate for 150 cc/min and to press

ACK

when ready.

With recommended sample gas pressure
(10–30 psig), the mass flow controller
will automatically control the flow at
150 ± 3 sccm. You may check the flow
by connecting a bubblometer at the
analyzer sample vent line to insure that
there is proper flow.

NOTE: Mass flow controller is preset to give 150 ± 3 sccm flow. However,

when a pneumatic valve does not open or the gas pressure drops to
a point where the mass flow controller is unable to deliver 150 sccm
flow, the analyzer will automatically shut down and display the
error message FLOW BELOW 20 SCCM. If this occurs, check the
sample/span gas and air pressure. Make sure that they are within
the recommended range.

14. After pressing

ACK

in step 13, the

system performs a purge-from-span,
and automatically continues to step 21.

FROM HERE GO TO STEP 21

VALVES: O FF

CONNECT THE SPAN GAS TANK
TO THE SPAN INLET

1

2

3

4

5

--- PRESS ACK W HEN READY ---

ACK

ESC

VALVES: O FF

CONNECT A SAM PLE GAS
TO THE SAM PLE PO RT

--- PRESS ACK W HEN READY ---

1

2

3

4

5

ACK

ESC

VALVES: PURGE-FROM-SPAN

FLOW = 150 CC/MIN LOW

1

2

3

4

5

CHECK THE FLOW RATE FOR
150 CC/MIN

--- PRESS ACK WHEN READY ---

ACK

ESC

VALVES: PURG E-FRO M -SPAN

PURG E IN PRO G RESS
TIM E REM AINING 00 : 01 : 00

PLEASE W AIT . . .

1

2

3

4

5

SKIP