Model 3060e – Teledyne 3060e - Ultra Trace oxygen analyzer User Manual

Page 60

background image

4-12

Model 3060E

Model 3060E

Model 3060E

Model 3060E

Model 3060E

Chapter 4

Chapter 4

Chapter 4

Chapter 4

Chapter 4

Teledyne Electronic Technologies

Analytical Instruments

2. After a few minutes, the screen displays

the options

FULL

or

QUICK

start mode.

Press

FULL

to select

FULL

START

.

NOTE: Holding down the orange key while pressing QUICK will allow you

to choose a cold start for the next restart.

3.

The screen instructs you to fill the
Ultra Trace sensor with electro-
lyte. Press

ACK

to continue or

ESC

to return to the previous screen.
This ensures that the sensor has
electrolyte.

4.

The next screen directs you to
select the background gas. After
selecting the gas, press

ACK

to

continue. Press

ESC

to return to

the previous screen.

5.

The system asks if you will use
the internal calibrator.

To use the internal calibrator
only, press

YES

and continue

on to step 6.

To set up for calibrations
using a span gas only, press

NO

and go to step 11.

To use both a span gas and the
internal calibrator, press

YES

and go to step 15.

VALVES: O FF

W ARNING: IF THIS IS THE FIRST
SETUP THE FULL STARTUP
IS REQ UIRED.

1

2

3

4

5

FULL

Q UICK

VALVES: OFF

CAREFULLY FILL THE SENSOR
W ITH ELECTROLYTE SO THAT THE
LEVEL IS BETW EEN FILL M ARKS
--- PRESS ACK W HEN READY ---

ESC

ACK

1

2

3

4

5

VALVES: OFF

SELECT THE BACKGROUND GAS

USING THE KNOB ON THE STATUS PANEL

ESC

ACK

1

2

3

4

5

VALVES: O FF

DO YO U W ANT TO USE THE
INTERNAL CALIBRATOR?

ESC

NO

1

2

3

4

5

YES