Forcing a Cold Start Without First Shutting Down
To force a
COLD
START
at the next power-up after the analyzer has been used,
follow the procedure outlined below.
1. From the Analyze mode main menu,
press the
STDBY
key.
2. The next menu will indicate that the
system is beginning to shut down, and
indicates the time remaining until the
next screen appears.
3. From the Stand-by mode, press
START
to continue.
4. The menu will prompt you to select the
background gas from the status panel.
Press
ESC
to go to the first system
menu.
5. Holding down the orange key while
pressing
QUICK
will allow you to
choose a
COLD
START
for the next
reboot.
VALVES: ANALYZE
FLO W = 150.0 CC/M IN
AUTORANGE: 10 PPM
1
2
3
4
5
RANGE
ALRM R
5.00
PPM 02
T=25C
C=9:59
SETUP
STDBY
CAL
VALVES: STANDBY
STANDBY M O DE - ALARM S DISABLED
START
1
2
3
4
5
VALVES: OFF
SELECT THE BACKGROUND GAS
USING THE KNOB ON THE STATUS PANEL
ESC
ACK
1
2
3
4
5
VALVES: OFF
WARNING: IF THIS IS THE FIRST
SETUP THE FULL STARTUP
IS REQUIRED.
1
2
3
4
5
FULL
QUICK
Hold down the third key while
pressing the second (QUICK) key.
VALVES: SHUTTING DO W N
TIM E REM AINING 00 : 02 : 00
PLEASE W AIT . . .