Forcing a Cold Start Without First Shutting Down

To force a

COLD

START

at the next power-up after the analyzer has been used,

follow the procedure outlined below.

1. From the Analyze mode main menu,

press the

STDBY

key.

2. The next menu will indicate that the

system is beginning to shut down, and
indicates the time remaining until the
next screen appears.

3. From the Stand-by mode, press

START

to continue.

4. The menu will prompt you to select the

background gas from the status panel.
Press

ESC

to go to the first system

menu.

5. Holding down the orange key while

pressing

QUICK

will allow you to

choose a

COLD

START

for the next

reboot.

VALVES: ANALYZE

FLO W = 150.0 CC/M IN

AUTORANGE: 10 PPM

1

2

3

4

5

RANGE

ALRM R

5.00

PPM 02

T=25C

C=9:59

SETUP

STDBY

CAL

VALVES: STANDBY

STANDBY M O DE - ALARM S DISABLED

START

1

2

3

4

5

VALVES: OFF

SELECT THE BACKGROUND GAS

USING THE KNOB ON THE STATUS PANEL

ESC

ACK

1

2

3

4

5

VALVES: OFF

WARNING: IF THIS IS THE FIRST
SETUP THE FULL STARTUP
IS REQUIRED.

1

2

3

4

5

FULL

QUICK

Hold down the third key while
pressing the second (QUICK) key.

VALVES: SHUTTING DO W N

TIM E REM AINING 00 : 02 : 00

PLEASE W AIT . . .

1

2

3

4

5