Model 3060e, Teledyne electronic technologies, Analytical instruments – Teledyne 3060e - Ultra Trace oxygen analyzer User Manual

Page 120: Table 5-8. high current failure

background image

5-12

Model 3060E

Model 3060E

Model 3060E

Model 3060E

Model 3060E

Chapter 5

Chapter 5

Chapter 5

Chapter 5

Chapter 5

Teledyne Electronic Technologies

Analytical Instruments

NO

YES

FAILURE

CORRECTED

YES

FAILURE

CORRECTED

HIGH CURRENT

FAILURE

(>30ppm)

CHECK FOR

AIR LEAKS

IN SAMPLE

LINE

YES

NO

NO

O-RINGS

PROPERLY INSTALLED

MOUNTING

BOLTS

TIGHT

PERFORM

INDICATED

SERVICE

(INSTALLATION)

CHECK SENSOR
MOUNTS

CHECK BYPASS VENT
IS SET TO ~ 1 L/MIN

SET SPARGER TO
MEDIUM

1

2

3

PRESS START KEY

TO RESET THE ANALYZER

TO FACTORY

SETTINGS

PROCEED

CALL

SERVICE

NO

FAILURE

CORRECTED

YES

REPLACE

ELECTROLYTE

PRESS START KEY

TO RESET THE ANALYZER

TO FACTORY

SETTINGS

Table 5-8. High Current Failure