8. The next screen prompts you to select

the auto settling slope for either zeroing
or spanning. To change the zero settling
slope, press

ZERO

.

9. Now use the

UP

and

DOWN

keys to

scroll the slope value you want to set,
and press

ACCPT

to enter this value into

the computer. (The system default for
the slope for zero settling is 0.005.)

NOTE: With the default zero slope of 0.005, a stable zero line can be

achieved for optimum accuracy at 1 ppb. However, if the accuracy
required is more than 5 ppb, a higher slope, such as 0.01 may be
selected. Figure 4-12 shows how changes in zero settling slope will
affect accuracy.

10. After pressing

ACCPT

in step 9, the next

screen brings you back to the auto-
settle parameter mode. To get to the
span settling slope menu, select

SLOPE

.

11. Press

SPAN

to continue.

12. Use the

UP

and

DOWN

keys to scroll the

slope value you want to set, and press

ACCPT

to enter this value into the

computer. (The system default for the
slope for span settling is 0.025.)

Note:

With the default span settling slope of 0.025, a very stable span
curve is obtained within 20–40 minutes. However, by increasing
the span settling value, calibration may be achieved much quicker
(10–20 minutes) but will be less accurate. Figure 4-13 shows how
accuracy is affected by changing span settling slope values.

VALVES: ANALYZE C=9:59

T=25C

SELECT THE AUTO SETTLE SLO PE
FO R EITHER ZERO O R SPAN

1

2

3

4

5

SPAN

ZERO

ESC

VALVES: ANALYZE C=9:59

T=25C

SET THE SLO PE FO R ZERO SETTLE.
A SM ALLER SLOPE M EANS A LO NGER
SETTLE TIM E BUT M O RE ACCURATE
CALIBRATIO N

SLO PE = 0.005

1

2

3

4

5

ESC

UP

DOW N

ACCPT

SLO PE

PERO D

VALVES: ANALYZE C=9:59

T=25C

SELECT AUTO -SETTLE PARAM ETER
TO M O DIFY - PERIO D, SLO PE
OR AUTO CALIBRATOR OFFSET

1

2

3

4

5

O FFST

ESC

VALVES: ANALYZE C=9:59

T=25C

SELECT THE AUTO SETTLE SLO PE
FO R EITHER ZERO O R SPAN

1

2

3

4

5

SPAN

ZERO

ESC

VALVES: ANALYZE C=9:59

T=25C

SET THE SLO PE FO R SPAN SETTLE.
A SM ALLER SLOPE M EANS A LO NGER
SETTLE TIM E BUT M O RE ACCURATE
CALIBRATIO N

SLO PE = 0.025

1

2

3

4

5

ESC

UP

DOW N

ACCPT