Toshiba Tohsiba Electromagnetic Flowmeter Converter L5232 User Manual

Page 98

background image

6F8A0917

- 97 -

Changing the pulse width

The following is an example to change the pulse width from 100ms to 250ms.

Switch operation

Display example

Description

H3: PLS WIDTH
100.0 ms

Select H3: PLS WIDTH from the setting item selection menu.
The currently set pulse width appears.

H3: PLS WIDTH
100.0 ms

The cursor appears.
Press

further to move the cursor to the desired digit to

change.

H3: PLS WIDTH
200.0 ms

250.0 ms

Change the number of the digit.
To change other digit, press

under this condition to move

the cursor to that digit and change the number there.

H3: PLS WIDTH
250.0 ms

The display blinks and the set value is displayed for

confirmation.

Press

MENU/ENT

again to write the data. Then the cursor

disappears and the display returns to the condition that the set

value is displayed.

H3: PLS WIDTH

Returns to the menu of setting item selection.

Note: If you try to set a value exceeding the settable range, *H. OVER SPEC. appears and the value returns to

the previously set value. Try to set a new value again.