Toshiba Tohsiba Electromagnetic Flowmeter Converter L5232 User Manual

Page 118

background image

6F8A0917

- 117 -

Changing the fixed output constant

The following is an example to change the fixed pulse output value from 50pps to 100pps. To
change the fixed current output value, select the function “L2: FIX. CURR.”

Switch operation

Display example

Description

L3: FIX. PULSE
0050 pps

Select L3: FIX. PULSE from the setting item selection menu.
The currently set fixed pulse output value appears.

L3: FIX. PULSE
0050 pps

The cursor appears.
Press

further to move the cursor to the desired digit to

change.

L3: FIX. PULSE
0150 pps

0100 pps

Change the number of the digit.
To change other digit, press

under this condition to move

the cursor to that digit and change the number there.

L3: FIX. PULSE
0100 pps

The display blinks and the set value is displayed for

confirmation.

Pres

MENU/ENT

again to write the data. Then the cursor

disappears and the display returns to the condition that the

setting value is displayed.

L3: FIX. PULSE

Returns to the menu of setting item selection.

Note: If you try to set a value outside of the settable range, the value will be forcibly set to the upper limit or the

lower limit of the settable range.