13.2.3 Flow rate indication is not stable

Use a power supply within the specified
range.

Ground the converter/detector with

Class D ground (100 or less ground
resistance) using IV wire of 5.5 mm²

minimum.

Design piping so that the detector pipe

is filled with fluid all the time.

YES

Design piping so that the fluid does not
carry bubbles.

NO

Isn't there large-current or high-

voltage cable or equipment near
the flowmeter?

YES

Install the flowmeter away from the

equipment producing electromagnetic or
electrostatic interference.

Contact your nearest Toshiba
representative.

Isn't the fluid in the detector pipe
carrying bubbles?

Is the detector pipe filled with

fluid?

Are converter and detector
grounded with Class D ground

(100 or less ground resistance)?

Is power supply voltage within
the specified range?

YES

NO

NO

YES

NO

YES

NO

START

Connect each cable securely to the

terminal block.

Aren't the cables slackened?

NO

YES