beautypg.com

Toshiba Tohsiba Electromagnetic Flowmeter Converter L5232 User Manual

Page 29

background image

6F8A0917

- 28 -

A

B

X

G

Y

E

Converter
terminal block

External ground
terminal

Twisting shields

Input signal cable

Twisting shields

Excitation cable

Detector terminal block

A

B

X

G

E

Y

FG

Ground terminal

Class D grounding
(Ground resistance 100 or less)

Figure 5.9 (b) Wiring between Detector and Converter

(when grounding for detector is

difficult)

Ground terminal

Conductive material pipe
Example: Metal, etc.

Non-conductive material pipe
Example: Resin pipe or metal pipe which

is internally coated with resin

Grounding wire

Ground terminal

Grounding wire

When piping material is non-conductive
Carry out class D grounding work
(grounding resistance 100 or less).

Class D ground
(Ground resistance
100 or less)

When piping material is conductive
Connect the grounding wires to both
ends of the piping flange.

Figure 5.10 Detector Grounding Method