Changing each setting for digital I/O

The following is an example to show how to change the function of Digital Output 1 from High
limit alarm output (2:HIGH ALM) to Fluid empty alarm (A:EMPTY ALM).
To change other setting values, select the following function numbers.

Digital Output 1

F1: DO1 FNCTN

Digital Output 2

F2: DO2 FNCTN

Digital Output 3

F3: DO3 FNCTN

Digital Output 4

F4: DO4 FNCTN

Digital Output alarm status 1

F5: DO1 ALM STS

Digital Output alarm status 2

F6: DO2 ALM STS

Digital Output alarm status 3

F7: DO3 ALM STS

Digital Output alarm status 4

F8: DO4 ALM STS

Digital Input 1

G1: DI1 FNCTN

Digital Input 2

G2: DI2 FNCTN

Digital Input control signal level 1

G3: DI1 DET LEV

Digital Input control signal level 2

G4: DI2 DET LEV

Switch operation

Display example

Description

F1: DO1 FNCTN
1:HIGH ALM

Select F1: DO1 FNCTN from the setting item selection menu.
The currently set function of Digital Output 1 appears.

F1: DO1 FNCTN
1:HIGH ALM

The cursor appears.

F1: DO1 FNCTN
A:EMPTY ALM

Select the output function A:EMPTY ALM.

F1: DO1 FNCTN
A:EMPTY ALM

The display blinks and the set value is displayed for

confirmation.

Press

MENU/ENT

again to write the data. Then the cursor

disappears and the display returns to the condition that the

setting value is displayed.

F1: DO1 FNCTN

Returns to the menu of setting item selection.