Toshiba Tohsiba Electromagnetic Flowmeter Converter L5232 User Manual
Page 93
6F8A0917
- 92 -
Changing each setting for digital I/O
The following is an example to show how to change the function of Digital Output 1 from High
limit alarm output (2:HIGH ALM) to Fluid empty alarm (A:EMPTY ALM).
To change other setting values, select the following function numbers.
Digital Output 1
F1: DO1 FNCTN
Digital Output 2
F2: DO2 FNCTN
Digital Output 3
F3: DO3 FNCTN
Digital Output 4
F4: DO4 FNCTN
Digital Output alarm status 1
F5: DO1 ALM STS
Digital Output alarm status 2
F6: DO2 ALM STS
Digital Output alarm status 3
F7: DO3 ALM STS
Digital Output alarm status 4
F8: DO4 ALM STS
Digital Input 1
G1: DI1 FNCTN
Digital Input 2
G2: DI2 FNCTN
Digital Input control signal level 1
G3: DI1 DET LEV
Digital Input control signal level 2
G4: DI2 DET LEV
Switch operation
Display example
Description
F1: DO1 FNCTN
1:HIGH ALM
Select F1: DO1 FNCTN from the setting item selection menu.
The currently set function of Digital Output 1 appears.
F1: DO1 FNCTN
1:HIGH ALM
The cursor appears.
F1: DO1 FNCTN
A:EMPTY ALM
Select the output function A:EMPTY ALM.
F1: DO1 FNCTN
A:EMPTY ALM
The display blinks and the set value is displayed for
confirmation.
Press
MENU/ENT
again to write the data. Then the cursor
disappears and the display returns to the condition that the
setting value is displayed.
F1: DO1 FNCTN
Returns to the menu of setting item selection.