Changing the fixed output function

The following is an example to change the fixed output function from Fixed value output OFF to
ON and then go on to set the current output value and the pulse output value. To change the current
output value or pulse output value independently, select the following function:

Fixed current output value

N2: FIX CURR

Fixed pulse output value

N3: FIX PULSE

Switch operation

Display example

Description

N1: FIXED OUT
2:OFF

Select N1: FIXED OUT from the setting item selection menu.
The currently set fixed value output On/Off setting appears.

N1: FIXED OUT
2:OFF

The cursor appears.

N1: FIXED OUT
1:ON

Select the fixed value output: ON.

N2: FIX. CURR.
10.0 mA

The display blinks and the set value is displayed for

conformation.

Press

MENU/ENT

again to write the data. The cursor

disappears and then the current output set value can be

changed.

N2: FIX. CURR.
10.0 mA

Move the cursor to the desired digit to change.

N2: FIX. CURR.
20.0 mA

Change the number of the digit.
To change other digit, press

under this condition to move

the cursor to the desired digit and change the number there.

N3: FIX. PULSE
100 PPS

The display blinks and the set value is displayed for
conformation.
Press

MENU/ENT

again to write the data. The cursor

disappears and then the pulse output set value can be changed.

N3: FIX. PULSE
100 PPS

The display blinks and the set value is displayed for

conformation.

Press

MENU/ENT

again to write the data. Then the cursor

disappears and the display returns to the condition that the

setting value is displayed.

N3: FIX. PULSE

Returns to the menu of setting item selection.

Note. If a value outside of the settable range is tried to set, the upper limit or lower limit value of the settable

range will be forcibly set.