Changing the preset output function

The following is an example to show how to change the preset output function from Output state
hold (1/HOLD) to One-shot pulse output (2:50ms PULSE).

Switch operation

Display example

Description

I2: PRESET FNC
1:HOLD

Select I2:PRESET FNC from the setting item selection menu.
The currently set preset count output function (HOLD in this
example) appears.

I2: PRESET FNC
1:HOLD

The cursor appears.

I2: PRESET FNC
2:50ms PULSE

Press

to select the setting code 1:50ms PULSE.

I2: PRESET FNC
2:50ms PULSE

The display blinks and the set value is displayed for

confirmation

Press

MENU/ENT

again to write the data. Then the cursor

disappears and the display returns to the condition that the

setting value is displayed.

I2: PRESET FNC

Returns to the menu of setting item selection.