Changing the meter size of the detector

The following is an example to change the meter size from 50mm to 10mm.

Switch operation

Display example

Description

A2: METER SIZE
50 mm

Select A2: METER SIZE from the setting item selection.
The currently set meter size (50mm in this example) appears.

A2: METER SIZE
50 mm

The cursor appears.

A2: METER SIZE
100 mm

Change the set value for meter size to 100mm.
The set value for meter size can be changed using

switch.

(Note 1)

A2: METER SIZE
100 mm

The display blinks and the set value is displayed for
confirmation.
Press

MENU/ENT

again to write the data. Then the cursor

disappears and the display returns to the condition that the
setting value is displayed.

A2: METER SIZE

Returns to the menu of setting item selection.

Note 1: The meter size is displayed cyclically as shown below:

* For partially filled pipes detectors (LF232*F) : 6~24inch (150~600mm )

Note 2: If the setting value for meter value is changed, the setting value by flow unit and the setting

value for exciting frequency of no concern may be rewritten.
After the meter size is set, be sure to check the setting values of Range, Count (Pulse) Rate
and Exciting Frequency.